Issue 02/2015 Art. No. 7001-0070 Operating Manual APT.line™ FED (E2) Drying and heating ovens with forced convection and enhanced timer functio
FED (E2) 02/2015 page 10/55 Keep safety labels complete and legible. Replace safety labels that are no longer legible. Contact BINDER Service fo
FED (E2) 02/2015 page 11/55 Symbol on the type plate Information CE conformity marking Electrical and electronic equipment manufactured / place
FED (E2) 02/2015 page 12/55 The drying and heating ovens FED do not dispose of any measures of explosion protection. DANGER Explosion hazard. D
FED (E2) 02/2015 page 13/55 Due to the special demands of the Medical Device Directive (MDD), these ovens are not qualified for sterilization of
FED (E2) 02/2015 page 14/55 2.1 Equipment overview FED (1) Display (2) Set-point value key (3) Selector keys (4) Time management key (5)
FED (E2) 02/2015 page 15/55 If you need to return the unit, please use the original packing and observe the guidelines for safe lifting and trans
FED (E2) 02/2015 page 16/55 3.4 Location of installation and ambient conditions Set up the drying and heating oven FED on an even and non-flamma
FED (E2) 02/2015 page 17/55 4. Installation 4.1 Electrical connection The drying and heating ovens FED are supplied ready for connection. The s
FED (E2) 02/2015 page 18/55 CAUTION The exhaust duct will become hot during operation. Danger of burning. ∅ Do NOT touch the exhaust duct duri
FED (E2) 02/2015 page 19/55 5.3 Air change Opening the air flap in the exhaust duct serves to adjust the air change. Without connecting a sucti
FED (E2) 02/2015 page 2/55 EC – declaration of conformity EG – KONFORMITÄTSERKLÄRUNG EC - DECLARATION OF CONFORMITY CE - DECLARATION DE CONFORM
FED (E2) 02/2015 page 20/55 5. Press button to return to normal display (actual value display) (automatically after approx. 30 seconds). Adjus
FED (E2) 02/2015 page 21/55 5. Set the desired fan speed with the buttons The fan speed can be set to a value between 0% and 100%. Wait 2 seco
FED (E2) 02/2015 page 22/55 6.3.1 Switching between Continuous operation and Timer operation Press the time management button . The controller
FED (E2) 02/2015 page 23/55 6.3.3 Setting the timer values 1. Press the time management button . The controller indicates its current time fu
FED (E2) 02/2015 page 24/55 6.4 User level settings By pressing down button in normal display (actual value display) for 5 sec, you enter the
FED (E2) 02/2015 page 25/55 6.4.1 Temperature unit change between degrees Celsius °C and degrees Fahrenheit °F If required, the temperature disp
FED (E2) 02/2015 page 26/55 1. Press down button for approx. 5 seconds. The display alternately shows “unit” and the temperature unit:
FED (E2) 02/2015 page 27/55 4. Set the required address with buttons . You can enter address values between 1 and 30. The set value is automat
FED (E2) 02/2015 page 28/55 6.4.5 Setting the interface mode and, if appropriate, the printer interval 1. Press down button for approx. 5 sec
FED (E2) 02/2015 page 29/55 8. Set the desired value from 0 to 255 with buttons . The printer intervals via the RS 422 interface can be set bet
FED (E2) 02/2015 page 3/55 Die oben beschriebenen Produkte sind konform mit folgenden harmonisierten Normen: The products described above are in
FED (E2) 02/2015 page 30/55 7. Temperature safety devices 7.1 Temperature safety device class 2 (DIN 12880) The temperature safety device class
FED (E2) 02/2015 page 31/55 1. Turn the control knob (7) of the safety device using a coin to its end-stop (position 10) (chamber protection).
FED (E2) 02/2015 page 32/55 If the safety device has taken over control (identifiable by the red alarm lamp (7a) lighting up and, in case of the
FED (E2) 02/2015 page 33/55 8. Options 8.1 Communication software APT-COM™ 3 DataControlSystem (option) The oven is regularly equipped with a s
FED (E2) 02/2015 page 34/55 8.4 Mostly gas-tight version (option for FED 53 and FED 115) With this option the oven is additionally sealed, so th
FED (E2) 02/2015 page 35/55 General information for safe handling of gas cylinders: • Store and use gas cylinders only in well ventilated areas
FED (E2) 02/2015 page 36/55 CAUTION Use of inappropriate plug. Danger of inflammation. To close the exhaust duct use the delivered plug only.
FED (E2) 02/2015 page 37/55 We recommend taking out a maintenance agreement. Please consult BINDER Service. BINDER telephone hotline: +49 (0) 74
FED (E2) 02/2015 page 38/55 We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough cleaning. Any corrosive dam
FED (E2) 02/2015 page 39/55 9.2.2 Decontamination The operator must ensure that proper decontamination is performed in case a contamination of t
FED (E2) 02/2015 page 4/55 Product registration
FED (E2) 02/2015 page 40/55 9.3 Sending the unit back to BINDER GmbH If you send a BINDER product to us for repair or any other reason, we will
FED (E2) 02/2015 page 41/55 10.2 Decommissioning Turn off units sizes 400 and 720 at the main power switch (10) and disconnect the unit from the
FED (E2) 02/2015 page 42/55 Prior to handing the unit over to a recycling company, it is the user’s responsibility that it is free from toxic, i
FED (E2) 02/2015 page 43/55 Certified companies disassemble waste BINDER equipment in primary substances for recycling according to directive 200
FED (E2) 02/2015 page 44/55 11. Troubleshooting Fault description Possible cause Required measures Temperature Set-point temperature is not reac
FED (E2) 02/2015 page 45/55 12. Technical description 12.1 Factory calibration and adjustment This unit was calibrated and adjusted in the fa
FED (E2) 02/2015 page 46/55 12.4 FED technical data Unit size 53 115 240 400 720 Exterior dimensions Width mm inch 634 24.96 834 32.83 1034 40.
FED (E2) 02/2015 page 47/55 Unit size 53 115 240 400 720 Ventilation data Air change at 70 °C / 158 °F x/h *** 29 19 17 11 at 150 °C / 302 °F x/
FED (E2) 02/2015 page 48/55 Unit size 53 115 240 400 720 Standard equipment (continued) Temperature safety device class 2 acc. to DIN 12880 with
FED (E2) 02/2015 page 49/55 12.6 Accessories and spare parts BINDER GmbH is responsible for the safety features of the unit only, provided skil
FED (E2) 02/2015 page 5/55 Content EC – declaration of conformity ...
FED (E2) 02/2015 page 50/55 13. Contamination clearance certificate 13.1 For units located outside North America and Central America Declaratio
FED (E2) 02/2015 page 51/55 3.3 Measures to be taken in case of skin contact or release into the atmosphere / Maßnahmen bei Personenkontakt oder
FED (E2) 02/2015 page 52/55 We herewith declare that the following measures have been taken / Wir erklären, dass folgende Maßnahmen getroffen wur
FED (E2) 02/2015 page 53/55 13.2 For units in North America and Central America Product Return Authorization Request Please complete this form a
FED (E2) 02/2015 page 54/55 Customer (End User) Decontamination Declaration Health and Hazard Safety declaration To protect the health of our em
FED (E2) 02/2015 page 55/55 4. Declaration of Decontamination For toxic, radioactive, biologically and chemically harmful or hazardous substanc
FED (E2) 02/2015 page 6/55 8. OPTIONS ...
FED (E2) 02/2015 page 7/55 Dear customer, For the correct operation of the drying and heating ovens with forced convection FED, it is important t
FED (E2) 02/2015 page 8/55 WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious (irrevers
FED (E2) 02/2015 page 9/55 Prohibition signs Do NOT touch Do NOT spray with water Information to be observed in order to ensure optimum fu